Description

PR 3112 Isolated Pt100 converter

Application

  • The 3112 temperature converter measures a standard 2-, 3- or 4-wire Pt100 temperature sensor, and provides an isolated analog voltage or current output.
  • High 3 port isolation provides surge suppression and protects the control system from transients and noise.
  • The 3112 can be mounted in the safe area or in Zone 2 / Division 2 areas.
  • Approved for marine applications.

Technical characteristics – 3112 Isolated Pt100 converter

  • Flexibly powered by 24 VDC (±30%) via power rail or connectors.
  • < 30 ms fast response time with simultaneous sensor error detection when selected.
  • Selectable 300 ms response time when signal dampening is needed.
  • Excellent conversion accuracy in all available ranges, better than 0.05% of span.
  • Meeting the NAMUR NE21 recommendations, the 3112 provides top measurement performance in harsh EMC environments.
  • The device meets the NAMUR NE43 standard defining out of range and sensor error output values.
  • A visible green LED indicates operational status of the unit and the input sensor.
  • All terminals are protected against overvoltage and polarity error.
  • High galvanic isolation of 2.5 kVAC.
  • Excellent signal/noise ratio of > 60 dB.

Mounting / installation / programming

  • Selectable DIP-settings for easy configuration of more than 1000 factory calibrated measurement ranges.
  • The narrow 6 mm housing allows up to 165 units to be mounted per meter of DIN rail, without any air gap between units.
  • Wide ambient temperature range of -25…+70°C.

Environmental Conditions

Operating temperature-25°C to +70°C
Storage temperature-40°C to +85°C
Calibration temperature20…28°C
Relative humidity< 95% RH (non-cond.)
Protection degreeIP20
Installation inPollution degree 2 & meas. / overvoltage cat. II

Mechanical specifications

Dimensions (HxWxD)113 x 6.1 x 115 mm
Weight approx.70 g
DIN rail typeDIN EN 60715/35 mm
Wire size0.13…2.5 mm2 / AWG 26…12 stranded wire
Screw terminal torque0.5 Nm
VibrationIEC 60068-2-6
2…25 Hz±1.6 mm
25…100 Hz±4 g

Common specifications – 3112 Isolated Pt100 converter

Supply

Supply voltage16.8…31.2 VDC
Max. required power0.70 W
Max. power dissipation0.70 W

Isolation voltage

Isolation voltage, test / working2.5 kVAC / 300 VAC (reinforced)
Zone 2 / Div. 2250 VAC

Response time

Response time (0…90%, 100…10%)< 30 ms / 300 ms (selectable)
AccuracyBetter than 0.05% of selected range
Signal / noise ratioMin. 60 dB
ProgrammingDIP-switches
Signal dynamics, input23 bit
Signal dynamics, output18 bit
EMC immunity influence< ±0.5% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst< ±1% of span
Incorrect DIP-switch setting identification0 V / 0 mA output; LED 0.5 s / 1 Hz

Input specifications – 3112 Isolated Pt100 converter

RTD input

Temperature range, Pt100-200…+850°C
Min. measurement range (span)10°C
Accuracy: the greater ofBetter than 0.05% of span or 0.1°C
Temperature coefficient: the greater of0.02°C/°C or ≤ ±0.01%/°C
Sensor current< 150 µA
Sensor cable resistance< 50 Ω per wire
Effect of sensor cable resistance (3-/4-wire)< 0.002 Ω / Ω
Sensor error detectionYes – selectable via DIP-switch
Broken sensor detection> 800 Ω
Shorted sensor detection< 18 Ω

Output specifications – 3112 Isolated Pt100 converter

Common output specifications

Updating time10 ms

Current output

Signal range0…23 mA
Programmable signal ranges0 / 4…20 mA
Sensor error indication (0…20 mA)0 mA or 23 mA / OFF
Sensor error indication (4…20 mA)3.5 mA or 23 mA / acc. to NAMUR NE43 or OFF
Load (@ current output)≤ 600 Ω
Load stability≤ 0.01% of span / 100 Ω
Current limitation @ low output load< 60 mA peak / < 4 mA average

Voltage output

Programmable signal ranges0/1…5 and 0/2…10 V
Sensor error indication0 V / 10% above the max. / none
Open output< 18 V

I.S. / Ex marking

ATEXII 3 G Ex ec IIC T4 Gc
IECExEx ec IIC T4 Gc
FM, USCl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, AEx nA IIC T4
FM, CACl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, Ex nA IIC T4
EAC Ex2Ex nA IIC T4 Gc X

Observed authority requirements

EMC2014/30/EU & UK SI 2016/1091
LVD2014/35/EU & UK SI 2016/1101
ATEX2014/34/EU & UK SI 2016/1107
RoHS2011/65/EU & UK SI 2012/3032
EACTR-CU 020/2011
EAC ExTR-CU 012/2011

Approvals -3112 Isolated Pt100 converter

ATEXKEMA 10ATEX0147 X
IECExKEM 10.0068X
UKEXDEKRA 21UKEX0055X
c FM usFM17US0004X / FM17CA0003X
c UL us, UL 61010-1E314307
EAC ExEAEU KZ 7500361.01.01.08756
DNV MarineTAA00001RW

 

Data sheet & Manuals- 3112 Isolated Pt100 converter


Data sheet

Manuals

 

Click here to visit Official product Details

Related Products

PR 4225 Universal fI ff Converter

$293.00 $338.00Price range: $293.00 through $338.00

5 in stock
Fast Delivery
PR 3105 Isolated converter

$185.00

1 in stock
Fast Delivery
PR 6437D HART 7 Temperature Transmitter

Available on backorder
Fast Delivery
PR 4114 Universal Transmitter

$255.00

7 Sold 10 in stock
Fast Delivery