Description

PR 3225 Universal Frequency Converter

Functional highlights

  • 3225 Universal Frequency Converter measures frequencies up to 100 kHz.
  • Active current output.
  • Buffered voltage output 10 VDC.
  • 2-point process calibration.
  • User-programmable trigger levels -0.05…6.5 V and sensor supply 5…17 V.
  • NAMUR sensor error detection.
  • Output relay with windows, setpoint and latch functionality.
  • Simulation of process value during commissioning / maintenance.
  • Fast response time, with simultaneous sensor error detection (PATENTED).
  • All terminals are over-voltage protected, polarity protected and short-circuit protected.

Technical highlights – PR 3225 Universal Frequency Converter

3225 Universal frequency converter

  • Accuracy < 0.06% / span.
  • Response time < 30 ms.
  • 2.5 kVAC, 3-port galvanic isolation.
  • Wide ambient temperature range -25…70°C.
  • NAMUR NE21, NE43.

Programming – PR 3225 Universal Frequency Converter

  • Easy configuration via DIP switches.
  • Factory calibrated in all selectable measurement ranges.
  • Configuration, monitoring, and diagnostics using PR 4500 detachable communication interfaces via the PR 4590 ConfigMate.
  • All programming can be password protected.
  • Scrolling help text in 7 languages.

Mounting – PR 3225 Universal Frequency Converter

  • Units can be mounted side by side, horizontally and vertically, without air gap on a standard DIN rail, even at 70°C ambient temperature.
  • Units can be supplied separately or installed on PR 9400 power rail.
  • The narrow 6.1 mm housing allows up to 163 units per meter.

Environmental Conditions

Operating temperature-25°C to +70°C
Storage temperature-40°C to +85°C
Calibration temperature20…28°C
Relative humidity< 95% RH (non-cond.)
Protection degreeIP20
Installation inPollution degree 2 & meas. / overvoltage cat. II

Mechanical specifications

Dimensions (HxWxD)113 x 6.1 x 115 mm
Weight approx.70 g
DIN rail typeDIN EN 60715/35 mm
Wire size0.13…2.5 mm2 / AWG 26…12 stranded wire
Screw terminal torque0.5 Nm

Common specifications

Supply

Supply voltage16.8…31.2 VDC
Fuse400 mA SB / 250 VAC
Max. required power≤ 1.2 W
Max. power dissipation0.65 W

Isolation voltage

Isolation voltage, test / working2.5 kVAC / 300 VAC (reinforced)

Response time

Response time (0…90%, 100…10%)≤ 30 ms

Auxiliary supplies

Sensor supply limitation23 mA, 5…17 V
Signal dynamics, output18 bit
Long term stability, current, 1yr / 5yr @ 25°C≤ 0.058% / ≤ 0.101%
Long term stability, voltage, 1yr / 5yr @ 25°C≤ 0.032% / ≤ 0.058%
AccuracySee manual for details
EMC immunity influence< ±0.5% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst< ±1% of span

Input specifications

Frequency input

Frequency range0.001 Hz to 100 kHz
Time range, time function10 μs to 999.9 s
Max. frequency / min. pulse width, with input filter ON75 Hz / 8 ms

Sensor specifications

Tacho, trig-level LOW / HIGH≤ -50 mV / ≥ +50 mV
NPN / PNP, trig-level LOW / HIGH≤ 4.0 V / ≥ 7.0 V
TTL, trig-level LOW / HIGH≤ 0.8 V / ≥ 2.0 V
S0, trig-level LOW / HIGH≤ 2.2 mA / ≥ 9.0 mA
NAMUR, trig-level LOW / HIGH≤ 1.2 / ≥ 2.1 mA

Special voltage / current input

User-programmable trig-levels-0.05…6.50 V 0.0…10.0 mA

Output specifications – PR 3225 Universal Frequency Converter

Current output

Signal range, active0…23 mA
Programmable signal ranges0 / 4…20 mA
Load (@ current output)≤ 600 Ω
Load stability≤ 0.001% of span / 100 Ω
Response time, programmable0.0…60.0 s
Sensor error indication0 / 3.5 / 23 mA / none
Current limit≤ 28 mA

Voltage output

Signal range≤ 11.5 VDC
Programmable signal ranges0…1; 0…5; 0…10; 0.2…1; 2…10 VDC
Load (@ voltage output)≥ 10 kΩ
Response time, programmable0.0…60.0 s

Relay output

Relay functionsSetpoint, Window and Latch
Hysteresis0…100%
ON and OFF delay0…3600 s
Power On delay0…9999 s
Sensor error reactionBreak / Make / Hold
Max. voltage250 VAC / 200 VDC
Max. current2 A
Max. AC power500 VA
Max. DC current, resistive load ≤ 30 VDC2 ADC
Max. DC current, resistive load > 30 VDCSee manual for details

Observed authority requirements

EMC2014/30/EU & UK SI 2016/1091
LVD2014/35/EU & UK SI 2016/1101
RoHS2011/65/EU & UK SI 2012/3032
ATEX2014/34/EU & UK SI 2016/1107

Approvals – PR 3225 Universal Frequency Converter

ATEXKEMA 10ATEX0147 X
IECExKEM 10.0068X
UKEXDEKRA 21UKEX0055X
c UL us, UL 61010-1E314307
CCC2020322310003554

 

Data sheet & Manuals –


Data sheet

Manuals

 

Additional information

Weight1.5 kg
Output

Analog, Relay

Related Products